- Turkish Journal of Chemistry
- Vol: 44 Issue: 5
- Determination of blood glucose parameter from human blood serum by using a quartz crystal microbalan...
Determination of blood glucose parameter from human blood serum by using a quartz crystal microbalance sensor coated with phthalocyanines compounds
Authors : Evin Sahin Sadik, Hamdi Melih Saraoglu, Ilke Gurol, Mehmet Ali Ebeoglu, Fatma Emel Kocak
Pages : 1293-1302
View : 14 | Download : 6
Publication Date : 9999-12-31
Article Type : Makaleler
Abstract :Determining the blood glucose level is important for the prevention and treatment of diabetes mellitus. We developed a sensor system using Quartz Crystal Ivlicrobalance (QCM) to determine the blood glucose level from human blood serum. This study consists of two experimental stages: artificial glucose/pure water solution tests and human blood serum tests. In the first stage of the study, the QCM sensor with the highest performance was identified using artificial glucose solution concentrations. In the second stage of the study, human blood serum measurements were performed using QCM to determine blood glucose levels. QCM sensors were coated with phthalocyanines (Pcs) by jet spray method. The blood glucose values of 96 volunteers, which ranged from 71 mg/dL to 329 mg/dL, were recorded. As a result of the study, human glucose values were determined with an average error of 3.25%.Keywords : Phthalocyanines, blood glucose, human blood serum, liquid sensing, quartz crystal microbalance